A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation

dc.contributor.authorZhao, Huaying
dc.contributor.authorGhirlando, Rodolfo
dc.contributor.authorAlfonso, Carlos
dc.contributor.authorArisaka, Fumio
dc.contributor.authorAttali, Ilan
dc.contributor.authorBain, David L.
dc.contributor.authorCölfen, Helmut
dc.contributor.authorRosenberg, Rose
dc.contributor.authorSchuck, Peter
dc.date.accessioned2015-08-05T13:39:59Z
dc.date.available2015-08-05T13:39:59Z
dc.date.issued2015eng
dc.description.abstractAnalytical ultracentrifugation (AUC) is a first principles based method to determine absolute sedimentation coefficients and buoyant molar masses of macromolecules and their complexes, reporting on their size and shape in free solution. The purpose of this multi-laboratory study was to establish the precision and accuracy of basic data dimensions in AUC and validate previously proposed calibration techniques. Three kits of AUC cell assemblies containing radial and temperature calibration tools and a bovine serum albumin (BSA) reference sample were shared among 67 laboratories, generating 129 comprehensive data sets. These allowed for an assessment of many parameters of instrument performance, including accuracy of the reported scan time after the start of centrifugation, the accuracy of the temperature calibration, and the accuracy of the radial magnification. The range of sedimentation coefficients obtained for BSA monomer in different instruments and using different optical systems was from 3.655 S to 4.949 S, with a mean and standard deviation of (4.304 ± 0.188) S (4.4%). After the combined application of correction factors derived from the external calibration references for elapsed time, scan velocity, temperature, and radial magnification, the range of s-values was reduced 7-fold with a mean of 4.325 S and a 6-fold reduced standard deviation of ± 0.030 S (0.7%). In addition, the large data set provided an opportunity to determine the instrument-to-instrument variation of the absolute radial positions reported in the scan files, the precision of photometric or refractometric signal magnitudes, and the precision of the calculated apparent molar mass of BSA monomer and the fraction of BSA dimers. These results highlight the necessity and effectiveness of independent calibration of basic AUC data dimensions for reliable quantitative studies.eng
dc.description.versionpublished
dc.identifier.doi10.1371/journal.pone.0126420eng
dc.identifier.pmid25997164eng
dc.identifier.ppn443174008
dc.identifier.urihttp://kops.uni-konstanz.de/handle/123456789/31528
dc.language.isoengeng
dc.rightsCC0 1.0 Universal
dc.rights.urihttp://creativecommons.org/publicdomain/zero/1.0/
dc.subject.ddc540eng
dc.titleA Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugationeng
dc.typeJOURNAL_ARTICLEeng
dspace.entity.typePublication
kops.citation.bibtex
@article{Zhao2015Multi-31528,
  year={2015},
  doi={10.1371/journal.pone.0126420},
  title={A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation},
  number={5},
  volume={10},
  journal={PLoS ONE},
  author={Zhao, Huaying and Ghirlando, Rodolfo and Alfonso, Carlos and Arisaka, Fumio and Attali, Ilan and Bain, David L. and Cölfen, Helmut and Rosenberg, Rose and Schuck, Peter},
  note={Article Number: e0126420}
}
kops.citation.iso690ZHAO, Huaying, Rodolfo GHIRLANDO, Carlos ALFONSO, Fumio ARISAKA, Ilan ATTALI, David L. BAIN, Helmut CÖLFEN, Rose ROSENBERG, Peter SCHUCK, 2015. A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation. In: PLoS ONE. 2015, 10(5), e0126420. eISSN 1932-6203. Available under: doi: 10.1371/journal.pone.0126420deu
kops.citation.iso690ZHAO, Huaying, Rodolfo GHIRLANDO, Carlos ALFONSO, Fumio ARISAKA, Ilan ATTALI, David L. BAIN, Helmut CÖLFEN, Rose ROSENBERG, Peter SCHUCK, 2015. A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation. In: PLoS ONE. 2015, 10(5), e0126420. eISSN 1932-6203. Available under: doi: 10.1371/journal.pone.0126420eng
kops.citation.rdf
<rdf:RDF
    xmlns:dcterms="http://purl.org/dc/terms/"
    xmlns:dc="http://purl.org/dc/elements/1.1/"
    xmlns:rdf="http://www.w3.org/1999/02/22-rdf-syntax-ns#"
    xmlns:bibo="http://purl.org/ontology/bibo/"
    xmlns:dspace="http://digital-repositories.org/ontologies/dspace/0.1.0#"
    xmlns:foaf="http://xmlns.com/foaf/0.1/"
    xmlns:void="http://rdfs.org/ns/void#"
    xmlns:xsd="http://www.w3.org/2001/XMLSchema#" > 
  <rdf:Description rdf:about="https://kops.uni-konstanz.de/server/rdf/resource/123456789/31528">
    <dc:contributor>Zhao, Huaying</dc:contributor>
    <dc:contributor>Attali, Ilan</dc:contributor>
    <dcterms:abstract xml:lang="eng">Analytical ultracentrifugation (AUC) is a first principles based method to determine absolute sedimentation coefficients and buoyant molar masses of macromolecules and their complexes, reporting on their size and shape in free solution. The purpose of this multi-laboratory study was to establish the precision and accuracy of basic data dimensions in AUC and validate previously proposed calibration techniques. Three kits of AUC cell assemblies containing radial and temperature calibration tools and a bovine serum albumin (BSA) reference sample were shared among 67 laboratories, generating 129 comprehensive data sets. These allowed for an assessment of many parameters of instrument performance, including accuracy of the reported scan time after the start of centrifugation, the accuracy of the temperature calibration, and the accuracy of the radial magnification. The range of sedimentation coefficients obtained for BSA monomer in different instruments and using different optical systems was from 3.655 S to 4.949 S, with a mean and standard deviation of (4.304 ± 0.188) S (4.4%). After the combined application of correction factors derived from the external calibration references for elapsed time, scan velocity, temperature, and radial magnification, the range of s-values was reduced 7-fold with a mean of 4.325 S and a 6-fold reduced standard deviation of ± 0.030 S (0.7%). In addition, the large data set provided an opportunity to determine the instrument-to-instrument variation of the absolute radial positions reported in the scan files, the precision of photometric or refractometric signal magnitudes, and the precision of the calculated apparent molar mass of BSA monomer and the fraction of BSA dimers. These results highlight the necessity and effectiveness of independent calibration of basic AUC data dimensions for reliable quantitative studies.</dcterms:abstract>
    <dc:creator>Rosenberg, Rose</dc:creator>
    <dc:contributor>Bain, David L.</dc:contributor>
    <void:sparqlEndpoint rdf:resource="http://localhost/fuseki/dspace/sparql"/>
    <dspace:hasBitstream rdf:resource="https://kops.uni-konstanz.de/bitstream/123456789/31528/3/Zhao_0-295818.pdf"/>
    <dc:creator>Attali, Ilan</dc:creator>
    <dc:contributor>Cölfen, Helmut</dc:contributor>
    <dc:contributor>Ghirlando, Rodolfo</dc:contributor>
    <dcterms:isPartOf rdf:resource="https://kops.uni-konstanz.de/server/rdf/resource/123456789/29"/>
    <dc:creator>Schuck, Peter</dc:creator>
    <dc:rights>CC0 1.0 Universal</dc:rights>
    <dc:date rdf:datatype="http://www.w3.org/2001/XMLSchema#dateTime">2015-08-05T13:39:59Z</dc:date>
    <dc:creator>Zhao, Huaying</dc:creator>
    <dc:creator>Bain, David L.</dc:creator>
    <dc:contributor>Arisaka, Fumio</dc:contributor>
    <dc:language>eng</dc:language>
    <dcterms:hasPart rdf:resource="https://kops.uni-konstanz.de/bitstream/123456789/31528/3/Zhao_0-295818.pdf"/>
    <dc:creator>Ghirlando, Rodolfo</dc:creator>
    <dc:creator>Cölfen, Helmut</dc:creator>
    <dc:creator>Alfonso, Carlos</dc:creator>
    <dcterms:rights rdf:resource="http://creativecommons.org/publicdomain/zero/1.0/"/>
    <dc:contributor>Rosenberg, Rose</dc:contributor>
    <bibo:uri rdf:resource="http://kops.uni-konstanz.de/handle/123456789/31528"/>
    <dc:contributor>Schuck, Peter</dc:contributor>
    <dspace:isPartOfCollection rdf:resource="https://kops.uni-konstanz.de/server/rdf/resource/123456789/29"/>
    <dc:creator>Arisaka, Fumio</dc:creator>
    <foaf:homepage rdf:resource="http://localhost:8080/"/>
    <dcterms:available rdf:datatype="http://www.w3.org/2001/XMLSchema#dateTime">2015-08-05T13:39:59Z</dcterms:available>
    <dcterms:issued>2015</dcterms:issued>
    <dcterms:title>A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation</dcterms:title>
    <dc:contributor>Alfonso, Carlos</dc:contributor>
  </rdf:Description>
</rdf:RDF>
kops.description.openAccessopenaccessgold
kops.flag.knbibliographytrue
kops.identifier.nbnurn:nbn:de:bsz:352-0-295818
kops.sourcefieldPLoS ONE. 2015, <b>10</b>(5), e0126420. eISSN 1932-6203. Available under: doi: 10.1371/journal.pone.0126420deu
kops.sourcefield.plainPLoS ONE. 2015, 10(5), e0126420. eISSN 1932-6203. Available under: doi: 10.1371/journal.pone.0126420deu
kops.sourcefield.plainPLoS ONE. 2015, 10(5), e0126420. eISSN 1932-6203. Available under: doi: 10.1371/journal.pone.0126420eng
relation.isAuthorOfPublication4bb493dd-f4c7-4ece-aab9-913000b58820
relation.isAuthorOfPublication65951f12-323c-4280-abc0-08d2bff8f787
relation.isAuthorOfPublication.latestForDiscovery4bb493dd-f4c7-4ece-aab9-913000b58820
source.bibliographicInfo.articleNumbere0126420eng
source.bibliographicInfo.issue5eng
source.bibliographicInfo.volume10eng
source.identifier.eissn1932-6203eng
source.periodicalTitlePLoS ONEeng
temp.internal.duplicates<p>Keine Dubletten gefunden. Letzte Überprüfung: 29.06.2015 14:05:59</p>deu

Dateien

Originalbündel

Gerade angezeigt 1 - 1 von 1
Vorschaubild nicht verfügbar
Name:
Zhao_0-295818.pdf
Größe:
1.64 MB
Format:
Adobe Portable Document Format
Zhao_0-295818.pdf
Zhao_0-295818.pdfGröße: 1.64 MBDownloads: 521

Lizenzbündel

Gerade angezeigt 1 - 1 von 1
Vorschaubild nicht verfügbar
Name:
license.txt
Größe:
3.59 KB
Format:
Item-specific license agreed upon to submission
Beschreibung:
license.txt
license.txtGröße: 3.59 KBDownloads: 0