Publikation:

A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation

Lade...
Vorschaubild

Dateien

Zhao_0-295818.pdf
Zhao_0-295818.pdfGröße: 1.64 MBDownloads: 447

Datum

2015

Autor:innen

Zhao, Huaying
Ghirlando, Rodolfo
Alfonso, Carlos
Arisaka, Fumio
Attali, Ilan
Bain, David L.
Schuck, Peter

Herausgeber:innen

Kontakt

ISSN der Zeitschrift

Electronic ISSN

ISBN

Bibliografische Daten

Verlag

Schriftenreihe

Auflagebezeichnung

ArXiv-ID

Internationale Patentnummer

Link zur Lizenz

Angaben zur Forschungsförderung

Projekt

Open Access-Veröffentlichung
Open Access Gold
Core Facility der Universität Konstanz

Gesperrt bis

Titel in einer weiteren Sprache

Publikationstyp
Zeitschriftenartikel
Publikationsstatus
Published

Erschienen in

PLoS ONE. 2015, 10(5), e0126420. eISSN 1932-6203. Available under: doi: 10.1371/journal.pone.0126420

Zusammenfassung

Analytical ultracentrifugation (AUC) is a first principles based method to determine absolute sedimentation coefficients and buoyant molar masses of macromolecules and their complexes, reporting on their size and shape in free solution. The purpose of this multi-laboratory study was to establish the precision and accuracy of basic data dimensions in AUC and validate previously proposed calibration techniques. Three kits of AUC cell assemblies containing radial and temperature calibration tools and a bovine serum albumin (BSA) reference sample were shared among 67 laboratories, generating 129 comprehensive data sets. These allowed for an assessment of many parameters of instrument performance, including accuracy of the reported scan time after the start of centrifugation, the accuracy of the temperature calibration, and the accuracy of the radial magnification. The range of sedimentation coefficients obtained for BSA monomer in different instruments and using different optical systems was from 3.655 S to 4.949 S, with a mean and standard deviation of (4.304 ± 0.188) S (4.4%). After the combined application of correction factors derived from the external calibration references for elapsed time, scan velocity, temperature, and radial magnification, the range of s-values was reduced 7-fold with a mean of 4.325 S and a 6-fold reduced standard deviation of ± 0.030 S (0.7%). In addition, the large data set provided an opportunity to determine the instrument-to-instrument variation of the absolute radial positions reported in the scan files, the precision of photometric or refractometric signal magnitudes, and the precision of the calculated apparent molar mass of BSA monomer and the fraction of BSA dimers. These results highlight the necessity and effectiveness of independent calibration of basic AUC data dimensions for reliable quantitative studies.

Zusammenfassung in einer weiteren Sprache

Fachgebiet (DDC)
540 Chemie

Schlagwörter

Konferenz

Rezension
undefined / . - undefined, undefined

Forschungsvorhaben

Organisationseinheiten

Zeitschriftenheft

Zugehörige Datensätze in KOPS

Zitieren

ISO 690ZHAO, Huaying, Rodolfo GHIRLANDO, Carlos ALFONSO, Fumio ARISAKA, Ilan ATTALI, David L. BAIN, Helmut CÖLFEN, Rose ROSENBERG, Peter SCHUCK, 2015. A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation. In: PLoS ONE. 2015, 10(5), e0126420. eISSN 1932-6203. Available under: doi: 10.1371/journal.pone.0126420
BibTex
@article{Zhao2015Multi-31528,
  year={2015},
  doi={10.1371/journal.pone.0126420},
  title={A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation},
  number={5},
  volume={10},
  journal={PLoS ONE},
  author={Zhao, Huaying and Ghirlando, Rodolfo and Alfonso, Carlos and Arisaka, Fumio and Attali, Ilan and Bain, David L. and Cölfen, Helmut and Rosenberg, Rose and Schuck, Peter},
  note={Article Number: e0126420}
}
RDF
<rdf:RDF
    xmlns:dcterms="http://purl.org/dc/terms/"
    xmlns:dc="http://purl.org/dc/elements/1.1/"
    xmlns:rdf="http://www.w3.org/1999/02/22-rdf-syntax-ns#"
    xmlns:bibo="http://purl.org/ontology/bibo/"
    xmlns:dspace="http://digital-repositories.org/ontologies/dspace/0.1.0#"
    xmlns:foaf="http://xmlns.com/foaf/0.1/"
    xmlns:void="http://rdfs.org/ns/void#"
    xmlns:xsd="http://www.w3.org/2001/XMLSchema#" > 
  <rdf:Description rdf:about="https://kops.uni-konstanz.de/server/rdf/resource/123456789/31528">
    <dc:contributor>Zhao, Huaying</dc:contributor>
    <dc:contributor>Attali, Ilan</dc:contributor>
    <dcterms:abstract xml:lang="eng">Analytical ultracentrifugation (AUC) is a first principles based method to determine absolute sedimentation coefficients and buoyant molar masses of macromolecules and their complexes, reporting on their size and shape in free solution. The purpose of this multi-laboratory study was to establish the precision and accuracy of basic data dimensions in AUC and validate previously proposed calibration techniques. Three kits of AUC cell assemblies containing radial and temperature calibration tools and a bovine serum albumin (BSA) reference sample were shared among 67 laboratories, generating 129 comprehensive data sets. These allowed for an assessment of many parameters of instrument performance, including accuracy of the reported scan time after the start of centrifugation, the accuracy of the temperature calibration, and the accuracy of the radial magnification. The range of sedimentation coefficients obtained for BSA monomer in different instruments and using different optical systems was from 3.655 S to 4.949 S, with a mean and standard deviation of (4.304 ± 0.188) S (4.4%). After the combined application of correction factors derived from the external calibration references for elapsed time, scan velocity, temperature, and radial magnification, the range of s-values was reduced 7-fold with a mean of 4.325 S and a 6-fold reduced standard deviation of ± 0.030 S (0.7%). In addition, the large data set provided an opportunity to determine the instrument-to-instrument variation of the absolute radial positions reported in the scan files, the precision of photometric or refractometric signal magnitudes, and the precision of the calculated apparent molar mass of BSA monomer and the fraction of BSA dimers. These results highlight the necessity and effectiveness of independent calibration of basic AUC data dimensions for reliable quantitative studies.</dcterms:abstract>
    <dc:creator>Rosenberg, Rose</dc:creator>
    <dc:contributor>Bain, David L.</dc:contributor>
    <void:sparqlEndpoint rdf:resource="http://localhost/fuseki/dspace/sparql"/>
    <dspace:hasBitstream rdf:resource="https://kops.uni-konstanz.de/bitstream/123456789/31528/3/Zhao_0-295818.pdf"/>
    <dc:creator>Attali, Ilan</dc:creator>
    <dc:contributor>Cölfen, Helmut</dc:contributor>
    <dc:contributor>Ghirlando, Rodolfo</dc:contributor>
    <dcterms:isPartOf rdf:resource="https://kops.uni-konstanz.de/server/rdf/resource/123456789/29"/>
    <dc:creator>Schuck, Peter</dc:creator>
    <dc:rights>CC0 1.0 Universal</dc:rights>
    <dc:date rdf:datatype="http://www.w3.org/2001/XMLSchema#dateTime">2015-08-05T13:39:59Z</dc:date>
    <dc:creator>Zhao, Huaying</dc:creator>
    <dc:creator>Bain, David L.</dc:creator>
    <dc:contributor>Arisaka, Fumio</dc:contributor>
    <dc:language>eng</dc:language>
    <dcterms:hasPart rdf:resource="https://kops.uni-konstanz.de/bitstream/123456789/31528/3/Zhao_0-295818.pdf"/>
    <dc:creator>Ghirlando, Rodolfo</dc:creator>
    <dc:creator>Cölfen, Helmut</dc:creator>
    <dc:creator>Alfonso, Carlos</dc:creator>
    <dcterms:rights rdf:resource="http://creativecommons.org/publicdomain/zero/1.0/"/>
    <dc:contributor>Rosenberg, Rose</dc:contributor>
    <bibo:uri rdf:resource="http://kops.uni-konstanz.de/handle/123456789/31528"/>
    <dc:contributor>Schuck, Peter</dc:contributor>
    <dspace:isPartOfCollection rdf:resource="https://kops.uni-konstanz.de/server/rdf/resource/123456789/29"/>
    <dc:creator>Arisaka, Fumio</dc:creator>
    <foaf:homepage rdf:resource="http://localhost:8080/"/>
    <dcterms:available rdf:datatype="http://www.w3.org/2001/XMLSchema#dateTime">2015-08-05T13:39:59Z</dcterms:available>
    <dcterms:issued>2015</dcterms:issued>
    <dcterms:title>A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation</dcterms:title>
    <dc:contributor>Alfonso, Carlos</dc:contributor>
  </rdf:Description>
</rdf:RDF>

Interner Vermerk

xmlui.Submission.submit.DescribeStep.inputForms.label.kops_note_fromSubmitter