Publikation: A practical EPR dosimetry system for routine use in radiotherapy : uncertainty analysis of lithium formate dosimeters at the therapeutic dose level
Dateien
Datum
Autor:innen
Herausgeber:innen
ISSN der Zeitschrift
Electronic ISSN
ISBN
Bibliografische Daten
Verlag
Schriftenreihe
Auflagebezeichnung
URI (zitierfähiger Link)
DOI (zitierfähiger Link)
Internationale Patentnummer
Link zur Lizenz
Angaben zur Forschungsförderung
Projekt
Open Access-Veröffentlichung
Sammlungen
Core Facility der Universität Konstanz
Titel in einer weiteren Sprache
Publikationstyp
Publikationsstatus
Erschienen in
Zusammenfassung
In electron paramagnetic resonance (EPR) dosimetry, solid dosimeter materials such as alanine (AL) or, more recently, lithium formate monohydrate (LFM) are typically used. These materials offer high potential for applications in radiotherapy based on their favorable dosimetric properties. Nevertheless, EPR dosimetry is not widespread in the clinics. This work presents an uncertainty analysis of EPR dosimetry in the dose range from 1 to 70 Gy using a compact spectrometer and applying a practical procedure being suitable for routine use in radiotherapy. The performances of self-pressed LFM pellets and commercial AL pellets are compared side by side. All pellets had a diameter of 4 mm and a height of 2 mm (AL) or 4 mm (LFM). The mean pellet mass was 35.81 mg and 73.81 mg for AL and LFM, respectively. Before irradiation, the pellets were stored for at least 8 weeks at 34± 2 % relative humidity. For irradiation, the pellets were put inside an airtight capsule. In total, 25 pellets per material were examined. The pellets were irradiated at a temperature of 25 ± 2.5 (2σ) °C to doses of either 1, 5, 20, 50 or 70 Gy (five pellets per dose value and material) by a clinical 6 MV photon beam. Measurement uncertainties were obtained from five independent readouts per pellet within five weeks following irradiation using a benchtop EPR spectrometer. The measurement time of a single readout was restricted to 10 min per pellet. Dose values were derived from EPR signal amplitudes using a specifically developed spectral fitting procedure. Signal fading characteristics were analyzed and taken into account during evaluation. The relative dose uncertainties (1σ) for a single readout at doses ≥ 5 Gy are below 2.8% (AL) and 1.1% (LFM) but increase to 12.3% (AL) and 2.6% (LFM) at 1 Gy. By averaging five independent readouts, the uncertainties at 1 Gy decrease to 2.6% (AL) and 0.8% (LFM). In terms of dose uncertainty, the LFM pellets are superior to the commercial AL pellets owing to their narrower EPR spectrum and approximately doubled mass resulting in higher EPR signal intensities. In case of the LFM pellets, the EPR dosimetry system shows a high level of precision (< 3%) down to 1 Gy being preferable for applications in radiotherapy. The uncertainties can be further decreased by averaging multiple dose values from independent readouts.
Zusammenfassung in einer weiteren Sprache
Fachgebiet (DDC)
Schlagwörter
Konferenz
Rezension
Zitieren
ISO 690
HÖFEL, Sebastian, Michael STEHLE, Felix ZWICKER, Michael K. FIX, Malte DRESCHER, 2021. A practical EPR dosimetry system for routine use in radiotherapy : uncertainty analysis of lithium formate dosimeters at the therapeutic dose level. In: Physics in Medicine and Biology. Institute of Physics Publishing (IOP). 2021, 66(4), 045005. ISSN 0031-9155. eISSN 1361-6560. Available under: doi: 10.1088/1361-6560/ab9414BibTex
@article{Hofel2021-02-05pract-51933, year={2021}, doi={10.1088/1361-6560/ab9414}, title={A practical EPR dosimetry system for routine use in radiotherapy : uncertainty analysis of lithium formate dosimeters at the therapeutic dose level}, number={4}, volume={66}, issn={0031-9155}, journal={Physics in Medicine and Biology}, author={Höfel, Sebastian and Stehle, Michael and Zwicker, Felix and Fix, Michael K. and Drescher, Malte}, note={Article Number: 045005} }
RDF
<rdf:RDF xmlns:dcterms="http://purl.org/dc/terms/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:rdf="http://www.w3.org/1999/02/22-rdf-syntax-ns#" xmlns:bibo="http://purl.org/ontology/bibo/" xmlns:dspace="http://digital-repositories.org/ontologies/dspace/0.1.0#" xmlns:foaf="http://xmlns.com/foaf/0.1/" xmlns:void="http://rdfs.org/ns/void#" xmlns:xsd="http://www.w3.org/2001/XMLSchema#" > <rdf:Description rdf:about="https://kops.uni-konstanz.de/server/rdf/resource/123456789/51933"> <dc:date rdf:datatype="http://www.w3.org/2001/XMLSchema#dateTime">2020-11-26T14:35:06Z</dc:date> <dcterms:rights rdf:resource="http://creativecommons.org/licenses/by/3.0/"/> <dcterms:isPartOf rdf:resource="https://kops.uni-konstanz.de/server/rdf/resource/123456789/29"/> <dcterms:abstract xml:lang="eng">In electron paramagnetic resonance (EPR) dosimetry, solid dosimeter materials such as alanine (AL) or, more recently, lithium formate monohydrate (LFM) are typically used. These materials offer high potential for applications in radiotherapy based on their favorable dosimetric properties. Nevertheless, EPR dosimetry is not widespread in the clinics. This work presents an uncertainty analysis of EPR dosimetry in the dose range from 1 to 70 Gy using a compact spectrometer and applying a practical procedure being suitable for routine use in radiotherapy. The performances of self-pressed LFM pellets and commercial AL pellets are compared side by side. All pellets had a diameter of 4 mm and a height of 2 mm (AL) or 4 mm (LFM). The mean pellet mass was 35.81 mg and 73.81 mg for AL and LFM, respectively. Before irradiation, the pellets were stored for at least 8 weeks at 34± 2 % relative humidity. For irradiation, the pellets were put inside an airtight capsule. In total, 25 pellets per material were examined. The pellets were irradiated at a temperature of 25 ± 2.5 (2σ) °C to doses of either 1, 5, 20, 50 or 70 Gy (five pellets per dose value and material) by a clinical 6 MV photon beam. Measurement uncertainties were obtained from five independent readouts per pellet within five weeks following irradiation using a benchtop EPR spectrometer. The measurement time of a single readout was restricted to 10 min per pellet. Dose values were derived from EPR signal amplitudes using a specifically developed spectral fitting procedure. Signal fading characteristics were analyzed and taken into account during evaluation. The relative dose uncertainties (1σ) for a single readout at doses ≥ 5 Gy are below 2.8% (AL) and 1.1% (LFM) but increase to 12.3% (AL) and 2.6% (LFM) at 1 Gy. By averaging five independent readouts, the uncertainties at 1 Gy decrease to 2.6% (AL) and 0.8% (LFM). In terms of dose uncertainty, the LFM pellets are superior to the commercial AL pellets owing to their narrower EPR spectrum and approximately doubled mass resulting in higher EPR signal intensities. In case of the LFM pellets, the EPR dosimetry system shows a high level of precision (< 3%) down to 1 Gy being preferable for applications in radiotherapy. The uncertainties can be further decreased by averaging multiple dose values from independent readouts.</dcterms:abstract> <dc:contributor>Höfel, Sebastian</dc:contributor> <dc:contributor>Zwicker, Felix</dc:contributor> <dcterms:hasPart rdf:resource="https://kops.uni-konstanz.de/bitstream/123456789/51933/1/H%c3%b6fel_2-1tufrg6e1pp905.pdf"/> <dc:contributor>Fix, Michael K.</dc:contributor> <dspace:hasBitstream rdf:resource="https://kops.uni-konstanz.de/bitstream/123456789/51933/1/H%c3%b6fel_2-1tufrg6e1pp905.pdf"/> <dc:creator>Drescher, Malte</dc:creator> <dspace:isPartOfCollection rdf:resource="https://kops.uni-konstanz.de/server/rdf/resource/123456789/29"/> <dcterms:issued>2021-02-05</dcterms:issued> <void:sparqlEndpoint rdf:resource="http://localhost/fuseki/dspace/sparql"/> <dc:contributor>Stehle, Michael</dc:contributor> <dc:language>eng</dc:language> <dc:creator>Stehle, Michael</dc:creator> <dc:creator>Höfel, Sebastian</dc:creator> <dc:creator>Zwicker, Felix</dc:creator> <foaf:homepage rdf:resource="http://localhost:8080/"/> <dc:creator>Fix, Michael K.</dc:creator> <bibo:uri rdf:resource="https://kops.uni-konstanz.de/handle/123456789/51933"/> <dc:rights>Attribution 3.0 Unported</dc:rights> <dc:contributor>Drescher, Malte</dc:contributor> <dcterms:available rdf:datatype="http://www.w3.org/2001/XMLSchema#dateTime">2020-11-26T14:35:06Z</dcterms:available> <dcterms:title>A practical EPR dosimetry system for routine use in radiotherapy : uncertainty analysis of lithium formate dosimeters at the therapeutic dose level</dcterms:title> </rdf:Description> </rdf:RDF>